Microplate Reader and Washer are available for diagnostic laboratories and research institutes worldwide.
Microplate Reader MPR-001
Variables | Specification | Measurement System | 8-channel optical system | Measurement Range | 0.00-3.500 ABs | Photometric Accuracy | ±1% or ±0.001Abs | Repeatability | < ±0.5% | Stability | <0.005Abs | Linearity | <±1% | Reading Speed | <5 secs/96 holes (single wavelength) <7secs/96 holes (double wavelength) | Filters | Standard: 7 filters (405, 450, 492, 630nm) or other wavelength on request |
Microplate Reader MPR-002
Variables | Specification | Measurement System | 2-channel optical system | Measurement Range | 0.001-3.500Abs | Photometric Accuracy | ±1% or ±0.001Abs | Repeatability | <±0.5% | Stability | <0.005Abs | Linearity | < ±1% | Reading Speed | < 5 sec/24 holes (single wavelength) <7 secs/24 holes (double wavelength) | Filters | Standard:450nm, 492nm |
Microplate Reader MPR-003
Variables | Specification | Measurement System | 8-channel optical system | Measurement Range | 0.001-4.000Abs | Photometric Accuracy | ±1% or ± 0.001 Abs | Repeatability | < ± 0.3% | Stability | < 0.002 Abs | Linearity | < ± 0.5% | Reading Speed | < 5 secs/96 holes (single wavelength) < 7 secs/96 holes (double wavelength) | Filters | Standard: 7 filters (405, 450, 492, 630 nm) or other wavelength on request | Display | Touch-screen |
Microplate Washer MPW-001
Variables | Specification | Washing Head | 8 or 12 chaneels | Applicable Microplate | 96, 48 or 32 wells microplate | Washing Times Option | 1-9, presettable | Washing Rows Option | 1-12, presetable | Soak Time | 1-3600s | Residual Volume | <2µl/well |
| |